We are not the manufacturer, authorised distributor or authorised representative of the Keyence brand in Great Britain. The private labels and part numbers shown on this website are the property of their respective owners.
Model | LS-7001*1 | ||
Type | Standard | ||
No. of connectable sensor heads | 2 (fully compatible for all head types except monitor (M) models) | ||
Display | Measurement display | Main display: 7-segment red LED (Character height: 20.3 mm) Sub-display: 7-segment red LED (Character height: 9.9 mm) x 3 | |
Minimum display unit | 0.01 to 100 µm (7-level selectable) | ||
Display range | ±99.99999 to 9,999.9 mm (Linked to minimum display unit setting, mm/µm selectable) | ||
Measurement position monitor | 7-level display with a red LED | ||
Tolerance check output display | Green LED (GO), Red LED x 2 (HI, LO) | ||
Input terminal block | Synchronous 1, 2 input | Non-voltage input for OUT1 | |
Auto-zero 1, 2 input | |||
Reset 1, 2 input | |||
Statistics 1, 2 input | - | ||
Program selection input | Non-voltage input x 4 inputs | ||
Output terminal | Analogue voltage output | ±10 V x 2 outputs | |
Alarm output | NPN open-collector output (N.C.) | ||
Total judgment output | NPN open-collector output for OUT1 | ||
Comparator ready output | |||
Strobe 1, 2 output | |||
Connector I/O | SUB mode | Statistical processing output | - |
Function output | Selectable from focus, area check, and differential, NPN open-collector output x 2 outputs*2 | ||
5-level comparator output | NPN open-collector output for OUT2*2 | ||
Comparator ready output | |||
Strobe output | |||
BCD mode | BCD output | Measurement data output (Sign + 7 digits), OUT1/OUT2 selectable, NPN open-collector output*2 | |
BCD selection output | NPN open-collector output*2 | ||
BCD selection input | Non-voltage input*2 | ||
Statistical processing input | - | ||
Synchronous input | Non-voltage input for OUT2 | ||
Reset input | |||
Auto-zero input | |||
Interface | RS-232C | Measurement data output and control I/O, printer (Baud rate can be selected up to 115,200 bps.) | |
Video output | - | ||
Major functions | Simultaneous measurement, Area designation, Calculation, Averaging, Calibration, 16-program memory, Measurement modes, Auto-zero, Print-out, Abnormal value elimination, Transparent object measurement, Mutual interference prevention, etc. | ||
Rating | Power voltage | 24 VDC ±10 %*3 | |
Power consumption | 0.7 A max.*3 | ||
Environmental resistance | Enclosure rating | IP64 (Panel surface only) | |
Ambient temperature | 0 to +40 °C | ||
Relative humidity | 35 to 85 % RH (No condensation) | ||
Weight | Approx. 820 g | ||
*1 The rating of the NPN open-collector inside the terminal block is: 100 mA max. (40 V max.), residual voltage of 0.5 V max. The rating of the NPN open-collector inside the connector I/O is: 30 mA max. (30 V max.), residual voltage of 0.5 V max. The rating of non-voltage input is: ON voltage of 1 V max., OFF current of 0.6 mA max. *2 Either SUB mode or BCD mode can be selected. *3 AC power supply can be used when the LS-S11 (AC power supply stand) is connected. |